Medical Vacuum Plant
The lubricated rotary vane vacuum pump is the most popular vacuum system technology used in hospitals and surgery centers today.
- Motor Range
- 1.5 – 25.0 hp/pump
- System Capacity
- 7.0 – 504.0 SCFM @ 19" HG
- Receiver Sizes
- 60 – 240 gallons
- Configurations
- Duplex / Triplex / Quadplex
- Duty Cycle
- 100%
- Cooling
- Air cooled — no water required
Highlights
- Sized and specified to your facility's demand
- NFPA99 compliant supply, installation & commissioning
- 24/7 after-sales support nationwide
- Genuine spare parts stocked locally
System Description
The lubricated rotary vane vacuum pump is the most popular vacuum system technology used in hospitals and surgery centers today. These rugged, oil-recirculating pumps have a long life and can provide thousands of hours of reliable service. Most are air cooled and require relatively little maintenance. In the medical industry, this product is popular from 1 HP through 25 HP and is excellent when designed into systems with multiple units.
The high quality pumps we use are a direct drive design requiring no belts and are driven by TEFC motors. These pumps are capable of operating continuously through their full vacuum range up to 29.3” HG. We fill each pump with a high quality, extended life synthetic lubricant. No water is required for the operation of these units – the water consumption savings per year is substantial.
Our ICU Lubricated Rotary Vane systems always include by-pass piping, vacuum gauge, double valve drain system, and sight gauge. We only use U.L. approved control panels and A.S.M.E. certified receivers. All systems are pre-plumbed and pre-wired for single point connections. Our vacuum systems always meet or exceed NFPA99 codes and other required regulations.
If you do not see what you are looking for in the standard systems shown below, we can custom build to suit your specific needs – generally at the same price as a standard system.
Features & Advantages
Equipment Features:
- TEFC motors are standard
- 100% Duty Cycle
- One piece pump and motor assembly
- Sealed and filtered recirculating oil system
- Air cooled design requires no water for operation
System Advantages:
- Direct drive eliminates belts and gears to loosen or wear
- Compact design for small spaces and easy installation
- Simple design is easily serviced on-site
- Internal oil mist eliminator for 99.9% oil-free exhaust
- Long life, low maintenance and low overall operating costs
Models & Capacities
Select any model number to open the manufacturer's technical specification pdf.
Duplex Vertical Space Saver Lubricated Rotary Vane NFPA Compliant Medical Vacuum Systems
* System capacities are shown with reserve vacuum pump as redundant standby to comply with NFPA 99.
Form #3120

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 1.5 | 7.0 | 7.0 | 80 | V73M-DTB-3VS |
| 1.5 | 7.0 | 7.0 | 120 | V73M-DTB-4VS |
| 2.0 | 11.0 | 11.0 | 80 | V73M-DTC-3VS |
| 2.0 | 11.0 | 11.0 | 120 | V73M-DTC-4VS |
| 3.0 | 17.0 | 17.0 | 80 | V73M-DTD-3VS |
| 3.0 | 17.0 | 17.0 | 120 | V73M-DTD-4VS |
Form #3104

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 5.0 | 26.0 | 26.0 | 80 | V73M-DTE-3VS |
| 5.0 | 26.0 | 26.0 | 120 | V73M-DTE-4VS |
| 5.0 (HD) | 37.0 | 37.0 | 80 | V73M-DTR-3VS |
| 5.0 (HD) | 37.0 | 37.0 | 120 | V73M-DTR-4VS |
Form #3105

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 7.5 | 52.0 | 52.0 | 120 | V73M-DTF-4VS |
| 10.0 | 65.0 | 65.0 | 120 | V73M-DTH-4VS |
| 10.0 (HD) | 77.0 | 77.0 | 120 | V73M-DTG-4VS |
Duplex Horizontal Lubricated Rotary Vane NFPA Compliant Medical Vacuum Systems
* System capacities are shown with reserve vacuum pump as redundant standby to comply with NFPA 99.
Form #3106

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 1.5 | 7.0 | 7.0 | 60 | V73M-DTB-2HS |
| 1.5 | 7.0 | 7.0 | 80 | V73M-DTB-3HS |
| 2.0 | 11.0 | 11.0 | 60 | V73M-DTC-2HS |
| 2.0 | 11.0 | 11.0 | 80 | V73M-DTC-3HS |
| 3.0 | 17.0 | 17.0 | 60 | V73M-DTD-2HS |
| 3.0 | 17.0 | 17.0 | 80 | V73M-DTD-3HS |
Form #3107

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 5.0 | 26.0 | 26.0 | 120 | V73M-DTE-4HS |
| 5.0 | 26.0 | 26.0 | 200 | V73M-DTE-5HS |
| 5.0 | 26.0 | 26.0 | 240 | V73M-DTE-6HS |
| 5.0 (HD) | 37.0 | 37.0 | 120 | V73M-DTR-4HS |
| 5.0 (HD) | 37.0 | 37.0 | 200 | V73M-DTR-5HS |
| 5.0 (HD) | 37.0 | 37.0 | 240 | V73M-DTR-6HS |
Form #3108

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 7.5 | 52.0 | 52.0 | 120 | V73M-DTF-4HS |
| 7.5 | 52.0 | 52.0 | 200 | V73M-DTF-5HS |
| 7.5 | 52.0 | 52.0 | 240 | V73M-DTF-6HS |
| 10.0 | 65.0 | 65.0 | 240 | V73M-DTH-6HS |
| 10.0 (HD) | 77.0 | 77.0 | 120 | V73M-DTS-4HS |
| 10.0 (HD) | 77.0 | 77.0 | 200 | V73M-DTS-5HS |
| 10.0 (HD) | 77.0 | 77.0 | 240 | V73M-DTS-6HS |
| 10.0 | 65.0 | 65.0 | 120 | V73M-DTH-4HS |
| 10.0 | 65.0 | 65.0 | 200 | V73M-DTH-5HS |
Duplex Stacked Lubricated Rotary Vane NFPA Compliant Medical Vacuum Systems
* System capacities are shown with reserve vacuum pump as redundant standby to comply with NFPA 99.
Form #3109

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 5.0 | 26.0 | 26.0 | 120 | V73M-DSE-4VS |
| 5.0 | 26.0 | 26.0 | 200 | V73M-DSE-5VS |
| 5.0 (HD) | 37.0 | 37.0 | 120 | V73M-DSR-4VS |
| 5.0 (HD) | 37.0 | 37.0 | 200 | V73M-DSR-5VS |
Form #3110

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 7.5 | 52.0 | 52.0 | 120 | V73M-DSF-4VS |
| 7.5 | 52.0 | 52.0 | 200 | V73M-DSF-5VS |
| 10.0 | 65.0 | 65.0 | 120 | V73M-DSH-4VS |
| 10.0 | 65.0 | 65.0 | 200 | V73M-DSH-5VS |
| 10.0 (HD) | 77.0 | 77.0 | 120 | V73M-DSG-4VS |
| 10.0 (HD) | 77.0 | 77.0 | 200 | V73M-DSG-5VS |
Form #3111

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 15.0 | 111.0 | 111.0 | 120 | V73M-DSJ-4VS |
| 15.0 | 111.0 | 111.0 | 200 | V73M-DSJ-5VS |
| 20.0 | 137.0 | 137.0 | 120 | V73M-DSK-4VS |
| 20.0 | 137.0 | 137.0 | 200 | V73M-DSK-5VS |
| 25.0 | 168.0 | 168.0 | 120 | V73M-DSL-4VS |
| 25.0 | 168.0 | 168.0 | 200 | V73M-DSL-5VS |
Triplex Horizontal Lubricated Rotary Vane NFPA Compliant Medical Vacuum Systems
Form #3112

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 5.0 (HD) | 37.0 | 74.0 | 240 | V73M-TTR-6HS |
| 5.0 | 26.0 | 52.0 | 120 | V73M-TTE-4HS |
| 5.0 | 26.0 | 52.0 | 200 | V73M-TTE-5HS |
| 5.0 | 26.0 | 52.0 | 240 | V73M-TTE-6HS |
| 5.0 (HD) | 37.0 | 74.0 | 120 | V73M-TTR-4HS |
| 5.0 (HD) | 37.0 | 74.0 | 200 | V73M-TTR-5HS |
Form #3113

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 7.5 | 52.0 | 104.0 | 120 | V73M-TTF-4HS |
| 7.5 | 52.0 | 104.0 | 200 | V73M-TTF-5HS |
| 7.5 | 52.0 | 104.0 | 240 | V73M-TTF-6HS |
| 10.0 | 65.0 | 130.0 | 120 | V73M-TTH-4HS |
| 10.0 | 65.0 | 130.0 | 200 | V73M-TTH-5HS |
| 10.0 | 65.0 | 130.0 | 240 | V73M-TTH-6HS |
| 10.0 (HD) | 77.0 | 154.0 | 120 | V73M-TTS-4HS |
| 10.0 (HD) | 77.0 | 154.0 | 200 | V73M-TTS-5HS |
| 10.0 (HD) | 77.0 | 154.0 | 240 | V73M-TTS-6HS |
Triplex Stacked Lubricated Rotary Vane NFPA Compliant Medical Vacuum Systems
* System capacities are shown with reserve vacuum pump as redundant standby to comply with NFPA 99.
Form #3114

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 5.0 | 26.0 | 52.0 | 120 | V73M-TVE-4VS |
| 5.0 | 26.0 | 52.0 | 200 | V73M-TVE-5VS |
| 5.0 (HD) | 37.0 | 74.0 | 120 | V73M-TVR-4VS |
| 5.0 (HD) | 37.0 | 74.0 | 200 | V73M-TVR-5VS |
Form #3115

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 7.5 | 52.0 | 104.0 | 120 | V73M-TVF-4VS |
| 7.5 | 52.0 | 104.0 | 200 | V73M-TVF-5VS |
| 10.0 | 65.0 | 130.0 | 120 | V73M-TVH-4VS |
| 10.0 | 65.0 | 130.0 | 200 | V73M-TVH-5VS |
| 10.0 (HD) | 77.0 | 154.0 | 120 | V73M-TVS-4VS |
| 10.0 (HD) | 77.0 | 154.0 | 200 | V73M-TVS-5VS |
Form #3116

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 15.0 | 111.0 | 222.0 | 200 | V73M-TSJ-5VS |
| 20.0 | 137.0 | 274.0 | 200 | V73M-TSK-5VS |
| 25.0 | 168.0 | 336.0 | 200 | V73M-TSL-5VS |
Quadplex Stacked Lubricated Rotary Vane NFPA Compliant Medical Vacuum Systems
* System capacities are shown with reserve vacuum pump as redundant standby to comply with NFPA 99.
Form #3117

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 5.0 (HD) | 37.0 | 111.0 | 120 | V73M-QSR-4VS |
| 5.0 (HD) | 37.0 | 111.0 | 200 | V73M-QSR-5VS |
| 5.0 | 26.0 | 78.0 | 120 | V73M-QSE-4VS |
| 5.0 | 26.0 | 78.0 | 200 | V73M-QSE-5VS |
Form #3118

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 7.5 | 52.0 | 156.0 | 120 | V73M-QSF-4VS |
| 7.5 | 52.0 | 156.0 | 200 | V73M-QSF-5VS |
| 10.0 | 65.0 | 195.0 | 120 | V73M-QSH-4VS |
| 10.0 | 65.0 | 195.0 | 200 | V73M-QSH-5VS |
| 10.0 (HD) | 77.0 | 231.0 | 120 | V73M-QSG-4VS |
| 10.0 (HD) | 77.0 | 231.0 | 200 | V73M-QSG-5VS |
Form #3119

| hp / pump | Capacity SCFM @ 19" HG | Receiver (gallons) | Model # | |
|---|---|---|---|---|
| per/pump | per/system* | |||
| 15.0 | 111.0 | 333.0 | 200 | V73M-QSJ-5VS |
| 20.0 | 137.0 | 411.0 | 200 | V73M-QSK-5VS |
| 25.0 | 168.0 | 504.0 | 200 | V73M-QSL-5VS |
Vibration Isolation
Proper vibration prevention is important to equipment efficiency, function and life expectancy. Not all vibration isolators are created equal and choosing the right ones for your equipment is instrumental.
Isolators are designed to create superior performance stability by absorbing vibrations and shock and by distributing the equipment load evenly. This prevents rocking, limits impact shock transmission and blocks sensitive components from low frequency shock and also high frequency vibrations present in the floor.
We use only the finest grade isolators chosen for your specific application, equipment load handling requirements and any NFPA99 or other regulations governing the equipment chosen for your facility.
Click on any link below to open a detailed product specification pdf:
- Neoprene Cork Pads
- Neoprene Cork Pads with Steel Plates
- Spring Mount Series Pads with California pre-approved Seismic Protection (OPA-0070)
Our Technical Department can help you choose which isolators are best suited for your application.

Medical Equipment Leasing & Finance
Leasing medical equipment makes perfect business sense. Leasing can provide the ability to purchase or replace new equipment, upgrade outdated equipment, add needed accessories, stay technologically advanced, streamline operations or grow your business without sacrificing cash reserves. Lease programs are tailor made and are designed specifically around your needs and business application.
Take advantage of these equipment leasing features and benefits:
- Conserve your capitol for other needed expenditures
- Conserve credit lines – a lease acts as an alternative credit line
- Allows for more efficient cash flow and cash reserves
- Saves money – a lease is paid for with before tax dollars
- Tax advantages – your lease may be a 100% tax-deductible operation expense
- Section 179 tax advantages may apply
- Leasing experts are available to guide you
*consult with your tax advisor for tax deductions, advantages and benefits.

Manufacturer Information
- Corporate Office
- 1954 Friendship Dr, Suite 103El Cajon, CA 92020
- Satellite Office
- Menifee, CA 92584
- Satellite Office
- Albuquerque, NM 87112
- Contact
- 1-800-743-8280
airandvac.com is a division of D & L Equipment Works, Inc. Specifications, model numbers and datasheets on this page are published by the manufacturer — view the original specification.